Overview
![]()
Andrew HR300B, Heat Regenerated Dehydrator with Alarms
Heat Regenerated Dehydrator, 19 in rack mountable, 20.7 kPa (3.0 psig), with low pressure, high pressure, high humidity, power and system failure alarms, 110-240 VAC: 48 Vdc
Specifications / Documents
- Alarm Type High:
- Humidity
- High pressure
- Low pressure
- Power failure
- System failure
- Cut In/Out Pressure Type: Continuous (default set to 20.7 kPa | 3.0 psig)
- Dehydrator Type: Automatically regenerating Desiccant
- Dew Point: Better than -45 °C (-49 °F) at 80% RH at +20 °C (+68 °F)
- Flow Rate, nominal: 5 SLPM
- High-Humidity Alarm: 10% ±2% RH, threshold
- High-Pressure Alarm Factory: set 8.5 psi
- Low-Pressure Alarm: Factory set 1.5 psi
- Mounting: Rack
- Port Count: 6
- Port Type: 3/8 in PE tube
- Power Cord: USA power cord
- Power-Fail Alarm: Loss of input power
- Rack Type EIA: 19 in, ETSI N3
- Rack Units: 3
- Regulated Output Pressure Type: Factory set, continuous
Dimensions
- Product Height: 132 mm, 5.197 in
- Product Width: 487 mm, 19.173 in
- Product Depth: 360 mm, 14.173 in
- Volume Capacity, maximum: 2 m³, 70.629 ft³
- Volume Capacity, minimum: 0.005 m³, 0.177 ft³
Electrical Specifications
- Alarm Contact Rating: Handles at least 0.5 A (ac or dc)
- Power Consumption, ac: 100 W
- Power Consumption, dc: 60 W
- Voltage: 110/240 Vac, 48/60 Vdc